Case with dental target containing a test object for internal and external quality control of dental radiology installations, both periapical and panoramic, with digital imaging systems. Delivery: 1 working day if ordered before 11:00 AM
529,00 € VAT included
Case enabling internal quality control of dental radiology
for digital periapical and panoramic radiographs.
If you have a panoramic and a retroalveolar camera of the same technology (digital),
only one phantom is needed.
For the retroalveolar (as standard): a stabilizing strap
(with integrated ergonomic system on the test object) combined with a mark engraved
on the latter allows all sensor and film formats to be accepted
and guarantees perfect reproducibility.
The DigiDent2 Phantom (digital image quality control) consists of:
a) Spatial resolution target
8 groups of line pairs:
2.50 – 2.80 – 3.10 – 3.70 – 4.40 – 5.00 – 5.60 – 6.30 line pairs per mm
The graduation is less than 20% from group to group.
b) Low contrast resolution test object
4 elements of diameters:
1mm – 1.5mm – 2.0mm – 2.5mm
c) Centering mark
Shipping costs included